Publication:
Evaluation of the performance of sysmex XN-3100 automated hematology analyzer regarding the sysmex XE-2100 and microscopic examination

dc.contributor.coauthorPalaoğlu, Kerim Erhan
dc.contributor.kuauthorİncir, Said
dc.contributor.kuprofileFaculty Member
dc.contributor.unitKoç University Hospital
dc.date.accessioned2024-11-09T12:26:36Z
dc.date.issued2021
dc.description.abstractObjectives: we performed a verification study of the Sysmex XN-3100 hematology analyzer in comparison with the XE-2100 according to the guidelines of the Clinical and Laboratory Standards Institute (CLSI) and the International Council for Standardization in Hematology (ICSH). Materials and methods: blood samples and quality control materials were used for precision. For comparison, we used the current XE-2100 as the comparative method and analyzed 540 blood samples. The Passing-Bablok and Bland-Altman tests were performed according to the CLSI EP09-A3 and a carryover study was performed according to the CLSI H26-A2 guidelines. The flagging performance of the two analyzers was compared, using two experienced laboratory technicians as the reference method. Results: the Sysmex XN-3100 demonstrated high levels of precision for most parameters. For the comparison analysis, all parameters, except for MCHC, monocytes and basophils were within the systematic error limits of desirable biological variability criterion (SeDBV). The carryover was less than 0.4% for all parameters. The flagging performance of the XN-3100 was satisfactory and the overall efficiency was high. Conclusions: the XN-3100 not only showed a strong correlation and agreement with the XE-2100 but also displayed a comparable analytical sensitivity, and increased specificity, which may result in an improved turnaround time and throughpu.
dc.description.fulltextYES
dc.description.indexedbyWoS
dc.description.indexedbyScopus
dc.description.issue1
dc.description.openaccessYES
dc.description.publisherscopeInternational
dc.description.sponsoredbyTubitakEuN/A
dc.description.sponsorshipN/A
dc.description.versionPublisher version
dc.description.volume46
dc.formatpdf
dc.identifier.doi10.1515/tjb-2020-0004
dc.identifier.eissn1303-829X
dc.identifier.embargoNO
dc.identifier.filenameinventorynoIR02792
dc.identifier.issn0250-4685
dc.identifier.linkhttps://doi.org/10.1515/tjb-2020-0004
dc.identifier.quartileN/A
dc.identifier.scopus2-s2.0-85103454776
dc.identifier.urihttps://hdl.handle.net/20.500.14288/1696
dc.identifier.wos625120300005
dc.keywordsBiological variation
dc.keywordsBlood cell count
dc.keywordsComparability
dc.keywordsHematology
dc.keywordsVerification
dc.languageEnglish
dc.publisherDe Gruyter
dc.relation.grantnoNA
dc.relation.urihttp://cdm21054.contentdm.oclc.org/cdm/ref/collection/IR/id/9443
dc.sourceTurkish Journal of Biochemistry
dc.subjectBiochemistry and molecular biology
dc.titleEvaluation of the performance of sysmex XN-3100 automated hematology analyzer regarding the sysmex XE-2100 and microscopic examination
dc.typeJournal Article
dspace.entity.typePublication
local.contributor.kuauthorİncir, Said

Files

Original bundle

Now showing 1 - 1 of 1
Thumbnail Image
Name:
9443.pdf
Size:
1 MB
Format:
Adobe Portable Document Format