Objective: Although sudden cardiac death is rare in children, an intracardiac defibrillator system is indicated in children with various types of cardiomyopathy, primary electrical diseases, and after surgical repair of congenital heart defects. The use of transvenous defibrillator lead systems is limited in pediatric patients because of a small body size and/or limited vascular access. Subcutaneous array leads combined with an abdominally placed generator can enable implantation. Method: This is a retrospective study of 13 patients who underwent subcutaneous defibrillator implantation between September 2010 and March 2015. The subcutaneous system was preferred because patients were not amenable to transvenous lead placement. Results: The median patient age was 4.1 years, and the median patient weight was 12.1 kg. Diagnoses of patients were long-QT syndrome in 6, aborted cardiac arrest with left ventricular non-compaction in 3, hypertrophic cardiomyopathy with sustained ventricular tachycardia in 3, and arrythmogenic right ventricular cardiomyopathy in 1. Revision of the subcutaneous lead was required in 5 patients 2-26 months after the implantation. Appropriate shocks were observed in three patients. Inappropriate shock and lead fractures were observed in one patient during the follow-up period. The failure of therapy was observed in one patient. There were no perioperative complications and no early or late deaths. Conclusion: Subcutaneous defibrillator systems are safe and effective in pediatric patients when the transvenous method is risky and contraindicated. Because the high growth rate in this population leads to lead failures, a close follow-up of this population is essential.

Source

Publisher

Turkish Society of Cardiology

Citation

item.page.haspartof

Source

Anatolian Journal of Cardiology

item.page.ispartofseries

item.page.edition

DOI

10.5152/AnatolJCardiol.2015.6589

item.page.datauri

item.page.link

Rights

CC BY-NC (Attribution-NonCommercial)

Copyrights Note

Creative Commons license

Except where otherwise noted, this item's license is described as CC BY-NC (Attribution-NonCommercial)

Endorsement

Review

Supplemented By

Referenced By

Related Patent

Related Goal

Item type:Goal, Access status: Open Access ,
02 - Zero Hunger
Hunger is the leading cause of death in the world. Our planet has provided us with tremendous resources, but unequal access and inefficient handling leaves millions of people malnourished. If we promote sustainable agriculture with modern technologies and fair distribution systems, we can sustain the whole world’s population and make sure that nobody will ever suffer from hunger again.
Item type:Goal, Access status: Open Access ,
03 - Good Health and Well-being
Over the last 15 years, the number of childhood deaths has been cut in half. This proves that it is possible to win the fight against almost every disease. Still, we are spending an astonishing amount of money and resources on treating illnesses that are surprisingly easy to prevent. The new goal for worldwide Good Health promotes healthy lifestyles, preventive measures and modern, efficient healthcare for everyone.
Google Scholar
Scholar'da Ara ↗
Altmetric
Dimensions
PlumX Metrikleri
BIP! Indicators