Publication: Thorough biomechanical analysis of arterial response to EasyEndo-Lite staple rotation: a simulation study in abaqus
Loading...
Program
KU-Authors
Organization Authors
Co-Authors
Cagan, Pinar
Kimiaei, Ali
Safaei, Seyedehtina
Kutlu, Cemal Asim
Date
Language
eng
Type
Embargo Status
No
Journal Title
Journal ISSN
Volume Title
Alternative Title
Abstract
Understanding instrument-tissue interaction is vital for safe surgery. We used Abaqus to simulate arterial response to EasyEndo-Lite Staple rotation, focusing on abrupt motion and strain-rate effects. Sudden angular rotation produced high strain rates and peak stresses in the clamped arterial segment, reaching 1.4 MPa at 15 degrees and similar to 3.7 MPa at 30 degrees. Rapid cessation of rotation increased forces by up to 28%, indicating elevated damage risk. The simulations provided detailed deformation and stress data across scenarios, demonstrating how optimizing instrument design and operating parameters could substantially reduce tissue trauma and improve surgical outcomes.
Source
Publisher
Taylor and Francis
Subject
Citation
item.page.haspartof
Source
Computer Methods in Biomechanics and Biomedical Engineering
item.page.ispartofseries
item.page.edition
DOI
10.1080/10255842.2026.2631137
item.page.datauri
item.page.link
Rights
N/A