Publication:
Bedside dynamic calculation of mechanical power: a validation study

dc.contributor.coauthorAsar, Sinan
dc.contributor.coauthorAcicbe, Özlem
dc.contributor.coauthorÇukurova, Zafer
dc.contributor.coauthorHergünsel, G. Oya
dc.contributor.coauthorCanan, Emral
dc.contributor.departmentN/A
dc.contributor.kuauthorÇakar, Nahit
dc.contributor.kuprofileFaculty Member
dc.contributor.schoolcollegeinstituteSchool of Medicine
dc.contributor.unitKoç University Hospital
dc.contributor.unitN/A
dc.contributor.yokid198906
dc.date.accessioned2024-11-09T23:18:42Z
dc.date.issued2020
dc.description.abstractPurpose: To develop an equation to calculate the bedside dynamic mechanical power (MPdyn) for modern ventilators using the Work of Breathing ventilator (WOBv) parameter. Materials and methods: We developed an equation based on mechanical power values, which is equal to WOBv x minute volume. To measure mechanical power with this equation forty adult patients, hospitalized with the diagnosis of Acute Respiratory Distress Syndrome and underwent invasive mechanical ventilation, were used. To be able compare our results with Gattinoni's standart mechanical power equation (MPstd) the contribution of the PEEP was included in our equation. Then results obtained from MP(dy)(n )and MPstd were compared using univariable regression and Bland-Altman analysis. This comparison was performed at different I:E ratios, PEEP levels and tidal volumes. Results: Analysis of the results for each condition showed that MPdyn, and MPstd equation correlated with R-2 >= 0.98. Additionally, there was no statistically significant difference between MPdyn and MPstd for patient power means were 0.04 J/min (p = .42) using Bland-Altman analysis. Conclusions: Physicians can easily calculate mechanical power by using MPdyn at the bedside of patients on volume control mode. (C) 2020 Elsevier Inc. All rights reserved.
dc.description.indexedbyWoS
dc.description.indexedbyScopus
dc.description.indexedbyPubMed
dc.description.openaccessNO
dc.description.publisherscopeInternational
dc.description.volume56
dc.identifier.doi10.1016/j.jcrc.2019.12.027
dc.identifier.eissn1557-8615
dc.identifier.issn0883-9441
dc.identifier.quartileQ2
dc.identifier.scopus2-s2.0-85077660854
dc.identifier.urihttp://dx.doi.org/10.1016/j.jcrc.2019.12.027
dc.identifier.urihttps://hdl.handle.net/20.500.14288/10428
dc.identifier.wos519555700027
dc.keywordsWork
dc.keywordsCare
dc.languageEnglish
dc.publisherW B Saunders Co-Elsevier Inc
dc.sourceJournal of Critical Care
dc.subjectCritical care medicine
dc.titleBedside dynamic calculation of mechanical power: a validation study
dc.typeJournal Article
dspace.entity.typePublication
local.contributor.authorid0000-0002-1302-9596
local.contributor.kuauthorÇakar, Nahit