Publication:
A short-term in vivo evaluation of the Istanbul heart left ventricular assist device in a pig model

Placeholder

School / College / Institute

Organizational Unit
Organizational Unit
Organizational Unit
SCHOOL OF MEDICINE
Upper Org Unit

Program

KU Authors

Co-Authors

Küçükaksu, Deniz Süha
Bakuy, Vedat
Arat, Nurcan
Erkasap, Pelin Çelikbilek
Aksoy, Emin

Publication Date

Language

Embargo Status

Journal Title

Journal ISSN

Volume Title

Alternative Title

Abstract

Objectives: a continuous-flow centrifugal blood pump system has been recently developed as an implantable left ventricular assist device for patients with endstage heart failure. The objective of this study was to evaluate the initial in vivo performance of a newly developed left ventricular assist device (iHeart or Istanbul heart; Manufacturing and Automation Research Center, Koc University, Istanbul, Turkey) in an acute setting using a pig model. Materials and Methods: three pigs (77, 83, 92 kg) received implants via a median sternotomy, with animals supported for up to 6 hours. An outflow cannula was anastomosed to the ascending aorta. Anticoagulation was applied by intravenous heparin administration. During the support period, pump performance was evaluated under several flow and operating conditions. All pigs were humanely sacrificied after the experiments, and organs were examined macroscopically and histopathologically. Results: flow rate ranged between 1.5 and 3.6 L/min with pump speeds of 1500 to 2800 revolutions/min and motor current of 0.6 to 1.3 A. Initial findings confirmed thatthe iHeart ventricular assist device had sufficient hydraulic performance to support the circulation. During the experimental period, plasma free hemoglobin levels were found to be within normalranges. Thrombus formation was not observed inside the pump in all experiments. Conclusions: the iHeart ventricular assist device demonstrated encouraging hemodynamic performance and good biocompatibility in the pig model for use as an implantable left ventricular assist device. Further acute in vivo studies will evaluate the short-term pump performance prior to chronic studies for long-term evaluation.

Source

Publisher

Middle East Society for Organ Transplantation

Subject

Transplantation

Citation

Has Part

Source

Experimental and Clinical Transplantation

Book Series Title

Edition

DOI

10.6002/ect.2019.0110

item.page.datauri

Link

Rights

Copyrights Note

Endorsement

Review

Supplemented By

Referenced By

0

Views

0

Downloads

View PlumX Details