Does the magnetic field of a magnetic stirrer in an optical aggregometer affect concurrent platelet aggregation?

Loading...
Thumbnail Image

Departments

Item type:Organizational Unit,

School / College / Institute

Item type:Organizational Unit,
SCHOOL OF MEDICINE
Upper Org Unit

Program

KU-Authors

Organization Authors

Co-Authors

Sağdilek, Engin

Çelebi, Gürbüz

Date

Language

Embargo Status

N/A

Journal Title

Journal ISSN

Volume Title

Alternative Title

Abstract

Platelets are subjected to extremely low frequency electromagnetic fields during standard aggregometry measurements owing to the use of a magnetic stir bar in the instrument. This study evaluates the effects of this magnetic field exposure on platelet aggregation by comparing the results obtained in a modified aggregometer. Blood samples from healthy volunteers were anticoagulated using citrate or heparin. Platelet-rich plasma (PRP) samples were prepared. A mechanical stirring device was attached to the aggregometer instead of the magnetic stir bar system. The PRP samples were stirred using a stirring rod tip that did not produce any magnetic fields in one channel of the aggregometer; in the other channel, a stirring rod carrying a small magnet at its tip was used. As a result, a magnetic field in the extremely low frequency range and in the amplitude range of 1.9–65 mT was applied to the platelets assigned to the channel where the magnetic stirring rod tip was used. Aggregation was induced using adenosine diphosphate (ADP), collagen, or epinephrine. The slopes, maximum aggregation values, and areas under the aggregation curves were compared between the magnetic and neutral stirring rod tip groups. For samples stirred with the magnetic stirring rod tip, a significant decrease was observed in 12 of the 14 parameters evaluated for aggregations induced with ADP or collagen compared to the neutral stirring rod tip, regardless of the method used for anticoagulation. This observation indicates that the magnetic stir bars used in standard aggregometry may significantly alter aggregation parameters and platelets may be possible targets of electromagnetic fields. 

Source

Publisher

Wiley

Citation

item.page.haspartof

Source

Bioelectromagnetics

item.page.ispartofseries

item.page.edition

DOI

10.1002/bem.21785

item.page.datauri

item.page.link

Rights

N/A

Copyrights Note

Rights and licensing

N/A

Endorsement

Review

Supplemented By

Referenced By

Related Patent

Related Goal

Google Scholar
Scholar'da Ara ↗
1
Görüntülenme
0
İndirme
Altmetric
Dimensions
PlumX Metrikleri
BIP! Indicators