Publication: Two cantilever based sytem for viscosity and density monitoring
Loading...
Program
Organization Authors
Co-Authors
Yaralıoğlu, Göksenin G.
Date
Language
Type
Embargo Status
N/A
Journal Title
Journal ISSN
Volume Title
Alternative Title
Abstract
Viscosity and density measurements in liquids in real-time is challenging. Previous MEMS based approaches use frequency sweeps for the purpose and those methods are slow and not real-time. We show that high precision viscosity and density measurements are possible using two cantilevers with different widths and by tracking their frequencies with a Phase-Locked-Loop in real-time.