Publication:
Simplified calculation of mechanical power for pressure controlled ventilation in COVID-19 ards patients

dc.contributor.coauthorAsar, Sinan
dc.contributor.coauthorAcicbe, Ozlem
dc.contributor.coauthorSabaz, Mehmet S.
dc.contributor.coauthorKucur Tulubas, Evrim
dc.contributor.coauthorHergunsel, Gulsum O.
dc.contributor.coauthorCukurova, Zafer
dc.contributor.coauthorCanan, Emral
dc.contributor.kuauthorÇakar, Nahit
dc.contributor.kuprofileFaculty Member
dc.contributor.schoolcollegeinstituteSchool of Medicine
dc.contributor.yokid198906
dc.date.accessioned2024-11-09T23:21:15Z
dc.date.issued2022
dc.description.abstractBackground: Mechanical power (MP) is a promising tool for guidance of lung protective ventilation. Different equations have been proposed to calculate MP in pressure control ventilation (PCV). The aim of this study is to introduce an easy to use MP equation MPpcv(m-simpl) and compare it to an equation proposed by Van der Meijden etal. (MPpcv) which considered as the reference equation in PCV. Methods: Ventilatory parameters of 206 COVID-19 ARDS patients recorded between 24-72 hours after admission to intensive care unit. The PCV data from these patients were retrospectively investigated. MP in PCV was calculated with a modified eq nation (MPpcv(m-simpl)) derived from the equation (MPpcv) ofVan der Meijden al: 0.098xRRxAVx(PEEP ; Delta P-insp insp - 1). The results from MPpcv(slope), MPpcv(simpl), and MPpcv(m-simpl), were compared to MP(pcv )at 15 cmH(2)O center dot s/L inspiratory resistance levels by univanable regression and Bland-Altman analysis. Results: Inspiratory resistance levels at 15 cmH(2)O s/L was found to be correlated between the power values calculated by MPpcv(simpl)/MPpcv(m-simpl),and the MPpcv(slope)/MPpcv based on univariable logistic regression (R-2 >98) analyses. In the comparison of all patients average MP values computed by the MPpcv(m-simpl) equation and the MPpcv reference equation. Bland-Altman analysis mean difference and p values at 15 cmH(2)O inspiratory resistance values (J/min) were found to be MPpcv(m-simpl) vs MPpcv =-0,04 (P=0.014); MPpcv(slope) vs. MPpcv =0.63 (P<0.0001); MPpcv(simpl) vs. MPpcv =0.64 J/min (P<0.0001), respectively. Conclusions: The results of this study confirmed that the MP(pcv(m-simpl) )equation can be used easily to calculate MP at bedside in pressure control ventilated COVID-19 ARDS patients.
dc.description.indexedbyWoS
dc.description.indexedbyScopus
dc.description.indexedbyPubMed
dc.description.issue44958
dc.description.openaccessNO
dc.description.publisherscopeInternational
dc.description.volume88
dc.identifier.doi10.23736/S0375-9393.21.15741-4
dc.identifier.eissn1827-1596
dc.identifier.issn0375-9393
dc.identifier.quartileQ2
dc.identifier.scopus2-s2.0-85125431428
dc.identifier.urihttp://dx.doi.org/10.23736/S0375-9393.21.15741-4
dc.identifier.urihttps://hdl.handle.net/20.500.14288/10860
dc.identifier.wos767346700008
dc.keywordsRespiratory distress syndrome
dc.keywordsVentilator-induced lung injury
dc.keywordsCOVID-19 mathematical-model
dc.languageEnglish
dc.publisherEdizioni Medico Scientifiche
dc.sourceMinerva Anestesiologica
dc.subjectAnesthesiology
dc.subjectCritical care medicine
dc.titleSimplified calculation of mechanical power for pressure controlled ventilation in COVID-19 ards patients
dc.typeJournal Article
dspace.entity.typePublication
local.contributor.authorid0000-0002-1302-9596
local.contributor.kuauthorÇakar, Nahit

Files