Determination of the biomechanical effect of an interspinous process device on implanted and adjacent lumbar spinal segments using a hybrid testing protocol: a finite-element study

Loading...
Thumbnail Image

Departments

Item type:Organizational Unit,
Item type:Organizational Unit,

School / College / Institute

Item type:Organizational Unit,
Item type:Organizational Unit,
SCHOOL OF MEDICINE
Upper Org Unit

Program

Organization Authors

Co-Authors

N/A

Date

Language

Embargo Status

N/A

Journal Title

Journal ISSN

Volume Title

Alternative Title

Abstract

Object: The authors evaluated the biomechanical effects of an interspinous process (ISP) device on kinematics and load sharing at the implanted and adjacent segments. Methods: A 3D finite-element (FE) model of the lumbar spine (L1-5) was developed and validated through comparison with published in vitro study data. Specifically, validation was achieved by a flexible (load-control) approach in 3 main planes under a pure moment of 10 Nm and a compressive follower load of 400 N. The ISP device was inserted between the L-3 and L-4 processes. Intact and implanted cases were simulated using the hybrid protocol in all motion directions. The resultant motion, facet load, and intradiscal pressure after implantation were investigated at the index and adjacent levels. In addition, stress at the bone-implant interface was predicted. Results: The hybrid approach, shown to be appropriate for adjacent-level investigations, predicted that the ISP device would decrease the range of motion, facet load, and intradiscal pressure at the index level relative to the corresponding values for the intact spine in extension. Specifically, the intradiscal pressure induced after implantation at adjacent segments increased by 39.7% and by 6.6% at L2-3 and L4-5, respectively. Similarly, facet loads at adjacent segments after implantation increased up to 60% relative to the loads in the intact case. Further, the stress at the bone-implant interface increased significantly. The influence of the ISP device on load sharing parameters in motion directions other than extension was negligible. Conclusions: Although ISP devices apply a distraction force on the processes and prevent further extension of the index segment, their implantation may cause changes in biomechanical parameters such as facet load, intradiscal pressure, and range of motion at adjacent levels in extension.

Source

Publisher

American Association of Neurological Surgeons

Citation

item.page.haspartof

Source

Journal Of Neurosurgery-Spine

item.page.ispartofseries

item.page.edition

DOI

10.3171/2014.12.SPINE14419

item.page.datauri

item.page.link

Rights

N/A

Copyrights Note

Rights and licensing

N/A

Endorsement

Review

Supplemented By

Referenced By

Related Patent

Related Goal

Item type:Goal, Access status: Open Access ,
03 - Good Health and Well-being
Over the last 15 years, the number of childhood deaths has been cut in half. This proves that it is possible to win the fight against almost every disease. Still, we are spending an astonishing amount of money and resources on treating illnesses that are surprisingly easy to prevent. The new goal for worldwide Good Health promotes healthy lifestyles, preventive measures and modern, efficient healthcare for everyone.
Google Scholar
Scholar'da Ara ↗
6
Görüntülenme
0
İndirme
Altmetric
Dimensions
PlumX Metrikleri
BIP! Indicators