Objective: For patient safety, maintaining hemodynamic stability during surgical procedures is critical. Dynamic indices [such as systolic pressure variation (SPV) and pulse pressure variation (PPV)], have recently seen an increase in use. Given the risks associated with such invasive techniques, there is growing interest in non-invasive monitoring methods-and plethysmographic waveform analysis. However, many such non-invasive methods involve intricate calculations or brand-specific monitors. This study introduces the simple systolic ratio (SSR), derived from pulse oximetry tracings, as a non-invasive method to assess fluid responsiveness. Methods: This prospective observational study included 25 adult patients whose SPV, PPV, and SSR values were collected at 30-min intervals during open abdominal surgery. The SSR was defined as the ratio of the tallest waveform to the shortest waveform within pulse tracings. The correlations among SSR, SPV, and PPV were analyzed. Additionally, anaesthesia specialists visually assessed pulse oximetry tracings to determine fluid responsiveness using the SSR method. Results: Strong correlations were observed between SSR and both SPV (r = 0.715, P < 0.001) and PPV (r = 0.702, P < 0.001). Receiver operator curve analysis identified optimal SSR thresholds for predicting fluid responsiveness at 1.47 for SPV and 1.50 for PPV. A survey of anaesthesia specialists using the SSR method to visually assess fluid responsiveness produced an accuracy rate of 83%. Conclusion: Based on the strong correlations it exhibits with traditional markers, SSR has great potential as a clinical tool, especially in resource-limited settings. However, further research is needed to establish its role, especially as it pertains to its universal applicability across monitoring devices. © 2024 by the Turkish Anesthesiology and Reanimation Association/Turkish Journal of Anaesthesiology & Reanimation is published by Galenos Publishing House.

Source

Publisher

Galenos Publishing House

Subject

Anesthesiology

Citation

Has Part

Source

Turkish Journal of Anaesthesiology and Reanimation

Book Series Title

Edition

DOI

10.4274/TJAR.2024.231452

item.page.datauri

Link

Rights

Copyrights Note

Endorsement

Review

Supplemented By

Referenced By

3

Views

3

Downloads

View PlumX Details