Publication: Paper integrated microfluidic contact lens for colorimetric glucose detection
Program
KU Authors
Co-Authors
İstif, Emin
Publication Date
Language
Type
Embargo Status
Journal Title
Journal ISSN
Volume Title
Alternative Title
Abstract
Contact lenses offer a simple, cost-effective, and non-invasive method for in situ real-time analysis of various biomarkers. Electro-chemical sensors are integrated into contact lenses for analysis of various biomarkers. However, they suffer from rigid electronic components and connections, leading to eye irritation and biomarker concentration deviation. Here, a flexible and microfluidic integrated paper-based contact lens for colorimetric analysis of glucose was implemented. Facilitating a three-dimensional (3D) printer for lens fabrication eliminates cumbersome cleanroom processes and provides a simple, batch compatible process. Due to the capillary force of the filter paper, the sample was routed to detection chambers inside microchannels, and it allowed further colorimetric detection. The paper-embedded microfluidic contact lens successfully detects glucose down to 2 mM within ∼10 s. The small dimension of the microfluidic system enables detection of glucose levels as low as 5 μl. The results show the potential of the presented approach to analyze glucose concentration in a rapid manner. It is demonstrated that the fabricated contact lens can successfully detect glucose levels of diabetic patients.
Source
Publisher
Royal Soc Chemistry
Subject
Biological marker, Biosensor, Wearable sensor
Citation
Has Part
Source
Sensors and Diagnostics
Book Series Title
Edition
DOI
10.1039/d4sd00135d