The safety and accuracy of the fluoroscopic imaging during proximal femoral fixation: a computerized 3D reappraisal of the joint penetration risk

Placeholder

Organizational Units

Program

KU Authors

Co-Authors

Advisor

Publication Date

2021

Language

English

Type

Journal Article

Journal Title

Journal ISSN

Volume Title

Abstract

Background: To assess the success of proximal cephalomedullary nailing operations for treating trochanteric fractures, surgeons utilize 2D fluoroscopy to observe the relative positions of the femoral head and the implant. One distance-based risk parameter, observed from the AP and Lateral projections, is the Tip-Surface Distance(TSD) that dictates how close to the outer cortex should the implant tip be residing to avoid post-surgical complications such as cut-out or joint penetration. In this study, the safety and the accuracy of the orthogonal fluoroscopic imaging were evaluated. Methods: A femoral head model was created and the risk zone was defined as a hemispherical shell of 5 mm thickness beneath the subchondral cortex, which should not be violated during screw insertion. The remaining hemisphere beneath the risk zone was designated as the safe zone. To assess the effect of head size, each simulation was conducted for 34, 47, and 60 mm diameter(Dfemur) femoral heads. The rate of safe zone violation was calculated for all possible screw endpoints with a TSD of at least 5 mm on fluoroscopic orthogonal views (TSDAP and TSDLat). Results: The minimum risk of joint penetration was achieved when the TSDAP/TSDLat ratio was 1. For Dfemur of 34 mm there was a risk of 91.7% of the safe zone violation when each TSDAP and TSDLat were 5 mm and 0% for 9 mm. For Dfemur of 47 mm, the risk was 92.2% for 5 mm and 0% for 11 mm. For Dfemur of 60 mm, the risk was 92.3% for 5 mm and 0% for 13 mm. Safety maps were constructed for all possible TSD combinations for 34, 47, and 60 mm femoral heads. Conclusions: Depending solely on the orthogonal fluoroscopic images is not a safe and accurate technique for assessing joint penetration risk during proximal femoral fixation due to the spherical geometry of the femoral head. The screw tip can lie completely outside of the femoral head even when it appears inside, in both orthogonal fluoroscopic views. Evidently, when using TSD, more stringent distance limits should be chosen, contrary to the recommended 5 mm limit. Our safety maps for TSD combinations may be used to check the security of the implantation. (C) 2020 Elsevier Ltd. All rights reserved.

Description

Source:

Injury-International Journal of The Care of The Injured

Publisher:

Elsevier Sci Ltd

Keywords:

Subject

Critical care medicine, Emergency medicine, Orthopedics, Surgery

Citation

Endorsement

Review

Supplemented By

Referenced By

Copy Rights Note

0

Views

0

Downloads

View PlumX Details